By learning the right way to maintain and care for your X-Ray apron, you will ensure adequate protection and extend the life of the apron. Below are orders and bans for proper care of the X-Ray apron
Check the apron for defects, cracks, bends and perforations. Place the X-Ray apron on a flat surface and visually check all seams as well as the outer and inner X-Ray apron covers for visible damage. Then check the straps and attachment devices to make sure they are in good condition. Finally, check the surface of the apron with your hands to locate any lumps, cracks, deflections or separation from the apron seams. If the condition of the apron seems to be suspect, it should be checked by radiography. It is recommended that the apron be examined fluoroscopically at least once a year.
Fluoroscopic check
- Put the X-Ray apron on the table
- Examine the entire X-Ray apron with a fluoroscope
Alternative control - when the fluoroscopic unit is not available.
- Take an X-Ray picture on the table
- Check for cracks or defects in the X-Ray material
Instructions on the suitability of aprons and protective accessories {link} can be found in the resources / user manuals tab
Tips for cleaning the apron
X-Ray aprons should always be cleaned after use.
- Clean by rubbing with a soft sponge, using cold water and a mild detergent based on active oxygen.
- Do not use preparations containing formaldehyde and chlorine compounds for disinfection.
- Do not iron
- Do not machine wash, autoclave or dry clean.
- Do not soak or immerse the X-ray apron in water.
- After cleaning, hang the apron in a vertical position on the intended hanger.
The X-Ray apron manufacturer's recommendations for the proper handling and storage of the apron must be strictly observed.
When the aprons are not in use, store them on hangers to prevent breaks in the protective material. If this is not possible, the X-Ray apron should be stored on a flat surface.
Aprons should never be folded or crinkled so as not to damage the internal protective material. "Cracks in the lead or lead-free lining may form in the overlap, reducing the life of the apron."
To prevent damage to the apron, avoid exposing the X-Ray apron to extremely high or low temperatures or direct sunlight.
Note: It is recommended that the manufacturer's instructions and regulations regarding the proper care and use of personal protective equipment be followed.